Ultrasonic calibration blocks

V1 Ultrasonic Calibration Block

V1 Ultrasonic Calibration Block

  • Application

V1 Calibration Block is designed for verification of the ultrasonic flaw detectors used in the course of welded joints quality control by the UT method. It is DSTU 4002-2000 (ISO 2400, DIN 54120, B.S. 2704) compliant.

V1 Calibration Block allows:

  • to determine the UT beam index and X-value with an error of ±0.5 mm
  • to determine the probe angle of the ultrasonic wave with an error of ±1 ang. degr.
  • to determine uncontrolled (“dead zone”) of the ultrasonic straight beam piezoelectric probes in steel
  • to determine resolution (in steel) of the ultrasonic straight beam piezoelectric probes
  • to determine the scan horizontal linearity of the ultrasonic flaw detectors
  • to adjust the scan speed and sensitivity of the straight beam, double crystal, angle beam piezoelectric transducers
  • Technical data

  • The Block is made from the low-carbon fine grain steel with small attenuation coefficient
  • ultrasonic longitudinal waves velocity in the Block material at a frequency of 5.0 MHz and a temperature of (20±5)ºC is (5920±30) m/s
  • ultrasonic longitudinal waves in the pad (polymethyl acrylate) of the Block material is at a frequency of 5.0 MHz and a temperature of (20±5)ºC is (2670±100) m/s
  • geometrical dimensions of the Block correspond to DSTU 4002-2000 (ISO 2400, DIN 54120, B. S. 2704)
  • the Block weight is up to 4.8 kg (approx.8.8 lbs).

About OKOndt GROUP

OKOndt GROUP represents a group of companies with more than 30 years of experience in research, development and production of non-destructive testing products for industrial customers around the world. The group's production companies are "Ultracon-Service" LLC and "Promprylad" LLC.

Need help?

CONTACT US

© 2008 - 2024 OKOndt GROUP. All Rights Reserved